Position the calibration weight on the weighing instrument
The difference between the weight value and the true mass is displayed, with ± sign.
External calibration Calibration record is printed, if adjustment was not performed and the process
Nom. + 10000.0 g was stopped by pressing (
Diff. - 0.3 g
--------------------
) Activate calibration/adjustment manually (press the ( key to cancel)
The calibration weight is displayed at the conclusion of calibration
-------------------- A GMP-compliant printout is generated
14.01.2002 13:00
Model CISL2
Ser.no. 12345678
Vers. 1.1007.12.1
BVers. 01-25-01
--------------------
External calibration
Nom. + 10000.0 g
Diff. - 0.3 g
External adjustment
Diff. + 0.0 g
--------------------
14.01.2002 13:02
Name:
--------------------
Unload the weighing instrument
C
o
m
b
ic
s 2
Co
m
b
ics 2